Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8: (1 year) CAL VER /5100,8842 INSTRUMENT: Fluke 10: (1 year) CAL VER /5100,8842 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 173 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 8842A STANDARD: Fluke 7405A-4207 AC Divider ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 7405A-4207 AC Divider 1.002 ASK- R N P F 1.003 ASK+ K 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP No warm-up time requirement. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 90%. 1.007 DISP Move Slide Switch to OFF position. 1.007 DISP 1.007 DISP Note: Erroneous results may occur if slide switch 1.007 DISP is not placed in OFF position before beginning tests. 1.008 HEAD {AC VOLTAGE TESTS} 1.009 DISP Move the Slide-Switch to the center position. 1.009 DISP Press SELECT for VAC. 1.010 HEAD AC VOLTAGE TESTS: {4V Range} 1.011 5100 4 0.010V 0.003U 50H 2W 2.001 5100 4 0.010V 0.003U 400H 2W 3.001 5100 0.000V S 2W 3.002 MEMI Enter UUT reading in volts AC: 3.003 MEME 3.004 MEMC 4 V +0.003U 4.001 5100 4 3.500V 0.105U 50H 2W 5.001 5100 4 3.500V 0.105U 400H 2W 6.001 HEAD AC VOLTAGE TESTS: {40V Range} 6.002 5100 40 35.00V 1.05U 50H 2W 7.001 5100 40 35.00V 1.05U 400H 2W 8.001 HEAD AC VOLTAGE TESTS: {400V Range} 8.002 5100 400 350.0V 10.5U 80H 2W 9.001 HEAD AC VOLTAGE TESTS: {600V Range} 9.002 5100 600 600V 20U 400H 2W 10.001 HEAD {DC VOLTAGE TESTS} 10.002 DISP Press SELECT for VDC. 10.002 DISP Press RANGE to select the 4V Range. 10.003 HEAD DC VOLTAGE TESTS: {4V Range} 10.004 5100 4 0.000V 0.002U 2W 11.001 HEAD DC VOLTAGE TESTS: {40V Range} 11.002 DISP Press RANGE to select the 40V Range. 11.003 5100 40 0.02V 0.01U 2W 12.001 HEAD DC VOLTAGE TESTS: {400V Range} 12.002 DISP Press RANGE to select the 400V Range. 12.003 5100 400 -0.2V 0.1U 2W 13.001 HEAD DC VOLTAGE TESTS: {4V Range} 13.002 DISP Press RANGE for 2s to select autorange. 13.003 5100 4 3.500V 0.055U 2W 14.001 HEAD DC VOLTAGE TESTS: {40V Range} 14.002 5100 40 -35.00V 0.54U 2W 15.001 HEAD DC VOLTAGE TESTS: {400V Range} 15.002 5100 400 100.0V 1.6U 2W 16.001 HEAD DC VOLTAGE TESTS: {600V Range} 16.002 5100 600 -600V 10U 2W 17.001 HEAD {CONTINUITY TEST} 17.002 DISP Move the Slide-Switch to the right position. 17.003 5100 2.000V 0.032U 2W 18.001 DISP Disconnect the 5500A from the UUT. 18.002 HEAD {RESISTANCE TESTS} 18.003 DISP Press SELECT for Ohms. 18.004 HEAD RESISTANCE TESTS: {400 Ohm Range} 18.005 DISP Short the UUT + and COM inputs. 18.006 MATH MEM1 = 0.0 18.007 MEMI Enter UUT reading in Ohms: 18.008 MEME 18.009 MEMC Z +0.2U 19.001 DISP Remove the short. 19.002 ASK- U 19.003 DISP Configure the 8842A as follows: 19.003 DISP [32] 8842A INPUT HI to 8842A SENSE HI 19.003 DISP [32] 8842A INPUT LO to 8842A SENSE LO 19.004 IEEE [@8842]G5[I] 19.005 MATH MEM1 = MEM - 1100 19.006 JMPF 19.009 19.007 DISP Select FRONT INPUTS on the 8842A. 19.008 JMP 19.004 19.009 DISP Connect the 5100B OUTPUT to the 8842A INPUT. 19.010 RESF 1Z S 2W 19.011 8842 1Z N 4W 19.012 MATH M[1] = MEM 19.013 RESF 100Z S 2W 19.014 8842 100Z N 4W 19.015 MATH M[2] = MEM 19.016 RESF 1kZ S 2W 19.017 8842 1kZ N 4W 19.018 MATH M[3] = MEM 19.019 RESF 10kZ S 2W 19.020 8842 10kZ N 4W 19.021 MATH M[4] = MEM 19.022 RESF * S 19.023 DISP Disconnect the leads from the 8842A and connect them to 19.023 DISP the UUT. 19.024 RESF 1Z S 2W 19.025 ASK+ U 19.026 MATH MEM = M[1] # 8842A 90 day accuracy, 20 Ohm Range 19.027 ACC Z .009% .0040U 19.028 MEMI Enter UUT reading in Ohms: 19.029 MEME 19.030 MEMC 400 Z 0.2U 20.001 ASK- U 20.002 RESF 100.0Z S 2W 20.003 ASK+ U 20.004 MATH MEM = M[2] # 8842A 90 day accuracy, 200 Ohm Range 20.005 ACC Z 0.007% 0.004U 20.006 MEMI Enter UUT reading in Ohms: 20.007 MEME 20.008 MEMC 400 Z 1.5% 0.2U 21.001 ASK- U 21.002 HEAD RESISTANCE TESTS: {4 kOhm Range} 21.003 RESF 1.000kZ S 2W 21.004 ASK+ U 21.005 MATH MEM = M[3] # 8842A 90 day accuracy, 2 kOhm Range 21.006 ACC kZ .005% .00003U 21.007 MEMI Enter UUT reading in kilohms: 21.008 MEME 21.009 MEMC 4 kZ 0.016U 22.001 ASK- U 22.002 HEAD RESISTANCE TESTS: {40 kOhm Range} 22.003 RESF 10.00kZ S 2W 22.004 ASK+ U 22.005 MATH MEM = M[4] # 8842A 90 day accuracy, 20 kOhm Range 22.006 ACC kZ .005% .0003U 22.007 MEMI Enter UUT reading in kilohms: 22.008 MEME 22.009 MEMC 40 kZ 1.5% 0.01U 23.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 23.002 RESF 400 100.0kZ 1.6U 2W 24.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 24.002 RESF 4 1.000MZ 1.5% 0.001U 2W 25.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 25.002 RESF 40 10.00MZ 1.9% 0.03U 2W 26.001 END